Research Article
Destabilization of the Comminuted Clavicle Shaft Fracture due to Breakage of the Titanium Locking Plate: Biomechanical Analysis
Grzegorz Szczęsny1* and Zbigniew Kowalewski2
1Department of Orthopedic Surgery and Traumatology, Medical University, Poland
2Department of Experimental Mechanics, IPPT, Polish Academy of Sciences, Poland
*Corresponding author: Grzegorz Szczęsny, Department of Orthopedic Surgery and Traumatology, Medical University, Warsaw, Poland
Published: 25 Sep, 2018
Cite this article as: Szczęsny G, Kowalewski Z.
Destabilization of the Comminuted
Clavicle Shaft Fracture due to
Breakage of the Titanium Locking Plate:
Biomechanical Analysis. Clin Surg.
2018; 3: 2125.
Abstract
Background: Implant breaking destabilizing the fracture may be caused by its structural defect,
inappropriate use, or massive overloads. We present the patient, who broke the plate stabilizing
comminuted clavicular fracture at the sixth postoperative week due to the, primarily reported,
simple loading of the limb that should not exceeded admissible limits.
Methods: Macroscopic examination excluded its manufacturing defect of an explant, but revealed
areas of brittle and fatigue fractures, as well as anterior and inferior bending of its edges. Mechanical
analysis proved sufficient material durability. Those findings clearly showed that the plate broke in
consequence of repetitive overloads exceeding its durability.
Patient, when acquitted with those results, confessed that the plate broke, while fishing. Moreover,
he reported an episode of alcohol overuse that could contribute to implant failure bringing the risk
of additional injury.
Results: Biomechanical analysis confirmed that mechanical loads throwing the spinner exceed
values adequate for plate breaking. Thus, the real circumstances of the complication seem to be
unveiled pointing out to facts that some patients disobey postoperative recommendations to avoid
overloading of stabilized extremity, thus participating actively in this type of complications, that
they dissemble the real circumstances of the complications, and that they may exhibit risky and
irrational behavior.
Conclusion: Basing on the presented case we concluded that in some cases plaster cast immobilization
supporting stabilized fracture may be beneficial precluding banned activities and preventing from
secondary injuries reducing the risk of postoperative complications.
Level of Evidence: Level III case control study
Keywords: Clavicle fracture; Implant failure; Mechanical overloads; Postoperative
recommendations disobey; Insubordinate patient
Introduction
Implant break off destabilizing the fracture regularly complicates orthopedic procedures. It may
be caused by structural defect of an implant, it’s inappropriate use, or massive overloads exceeding
its strength [1,2]. The first one is casuistic, as producer’s quality control effectively excludes defective
ones. The second may result from application of an implant dedicated to the bone carrying out much
smaller loads than stabilized one, but may also originate from repetitive implant bending adopting
it to the anatomical shape of the bone, thus reflecting the technical error during the procedure
[3]. The third one - when the healing processes do not support implant’s durability and excessive
loads are applied over the fracture gap [4,5]. Clinical observations point out that in some cases the
third reason occurs and implant failure originates from an excessive limb loads in non-immobilized
fractures in early or in un-united fractures in late stages of the healing process [6].
An analysis of the reason of implant breaking usually brings about troubles coming from
limited insight into the incident, as patients often dissemble the genuine circumstances that led
to the complication avoiding its responsibility. Especially when they do not follow postoperative
directions. Moreover, implant failure may also be used for clamor of compensation, bringing
about several troubles to the manufacturer, to the hospital, and to
the surgeon. Thus, recognizing the reason for implant breakings is
of value not only from medical, but also juridical and financial point
of view.
We present the patient, to whom the locking plate stabilizing
comminuted clavicular fracture broke at the sixth postoperative week
due to the primarily reported simple loading of the operated place
that should not exceed admissible limit. Since circumstances of the
incidence seemed to be ambiguous, material and mechanical analyses
of the implant were performed, and anamnesis was reevaluated
bringing about conclusions that we found important for other
patients subjected to the analogical treatment.
Figure 1
Figure 1
Preoperative x-ray revealing comminuted clavicle shaft fracture
with two intermediate bone fragments (arrows).
Figure 2
Material and Methods
A 33-years old, athletic male (105 kg body weight, 188 cm height)
was operated on due to comminuted fracture of the shaft of the left
clavicle (Robinson type II, B2; (Figure 1). Fracture was anatomically
reduced and stabilized with 2.8 mm Ti-6Al-4V, precontoured, angle
stable, locking plate (System 5.0, ChM, Poland) implanted on its
superior surface. Both intermediate bone fragments were additionally
fixed with titanium cortical screws (Figure 2). Postoperative
wound healed uneventfully and patient was discharged with limb
immobilized in shoulder brace for four weeks. After that, patient was
referred to rehabilitation, but was recommended to avoid massive
physical loads.
At the end of the sixth postoperative week, he reported a sudden
pain and noticeable click at the shoulder closing the trunk lid of the
car. Performed x-ray revealed broken plate and destabilized fracture
(Figure 3). Patient was re-operated and plate was replaced. No any
signs or symptoms of infection were noticed both macroscopically at
the operation field and in laboratory tests. Again, the postoperative
wound healed uneventfully, but patient was discharged with limb
immobilized in plaster cast for four weeks. Afterwards, he was
referred to rehabilitation.
Clinical controls and X-rays during the forthcoming 12 months
showed properly proceeding regenerative process and restoration of
the physiological limb function.
Figure 3
Figure 3
X-ray showing destabilized fracture and broken plate. Proximal
aspect of the clavicle rotated anteriorly, superior angulation of the plate
fragments.
Figure 4
Figure 4
Macrophotograph of the site of break. Dashed black arrow runs
from posterior to the anterior surface of the clavicle. Areas of brittle (matt
gray, white arrows) and fatigue (darker; black arrow) fracture.
Results
Analysis of an X-ray
Shoulder X-ray in PA position made on readmission showed
broken plate at the place of its hole. The fracture was destabilized.
Moreover, medial fragment was rotated anteriorly, and the clavicle
was angulated superiorly (Figure 3).
Macroscopic analysis of the broken plate
The explant was analyzed macroscopically excluding its
manufacturing defect. Under magnification, the areas of brittle and
fatigue fractures were identified (Figure 4). Interestingly, fatigue
fracture was localized on the inferior part of its anterior branch, and
brittle-on the posterior as well as inferior aspect of the anterior branch
that was, additionally, curved anteriorly and inferiorly (Figure 5).
Mechanical analysis of the plate
Mechanical properties of Ti-6Al-4V alloy, that is its ultimate
tensile strength of ca 960 MPa and yield point of ca 850 MPa, are well
known, as it is regularly used for implant’s manufacturing. Measuring
plate dimensions (length, thickness and width) and the length of the
lateral clavicle fragment (serving as a lever) allowed estimating values
of admissible loads that, when exceeded, resulted in the break.
Assuming an uniaxial load with driving force acting along the
plate, the estimated plastic flow starts at 10 kN, and plate break at 11,4
kN. But it could practically never occur. In vast majority of cases, the
clavicle moves in transverse plane forward and backward, bending
the plate at its hole and stretching its posterior branch, whereas the
anterior one serves as a hinge. Under those conditions, the load
of 580 N only is sufficient to break the plate, when acting on the
acromioclavicular joint from its posterior aspect. Angulation of the
antero-inferior edge of the plate supports this observation pointing
out to the directions of loads that resulted in the break (Figure 6).
Moreover, orienting the plate to the patient’s body it could also be
concluded that the break was caused by a combination of multiaxial
loads under stress conditions, resulting from simultaneous effect of
tension, bending and torque. It clearly shows that the patient has not
respected the post-operative recommendations to avoid overloads.
Carried out theoretical calculations were verified experimentally
subjecting the analogical plates to simple tension and bending. They
revealed that the permanent deformation begins at the axial load
of 7,5 kN, leading to its break at the force of 9,4 kN. But it could
never happen, as loads of such magnitudes are unlikely. Nevertheless,
simple bending with relatively small force (40 N), but acting over
an arm of 90 mm (the distance from the site of break to the clavicoacromial
joint), thus producing the moment of the force of 3,6 Nm,
are sufficient to initiate the permanent deformation.
In the presented case the most probable loads that resulted in
plate break off acted in the transverse plane, stretching the posterior
branch of the plate, whereas the anterior branch served as a hinge
and plate width as a lever. Under those conditions the moment of
the force that deforms permanently the plate was validated to be 12,8
Nm, and it corresponds to that theoretically estimated. Thus, both
theoretical and experimental investigations showed that loads of 160
N (for instance raising the weight of 16 kg from the ground) acting
over the plate in transverse plane over the shoulder from its posterior
aspect, is sufficient to deform it permanently.
Muscle attachments to the clavicle
Characteristic displacement of bone fragments observed on readmission
X-ray could be explained analyzing muscle attachments.
While the medial part of the clavicle is elevated by sterno-cleidomastoid
muscle, the lateral one is brought down by the weight of
the extremity and tension of the deltoid and sub-scapular muscles.
Rotation (anteversion) of the medial bone fragment could only be
explained by the strong constriction of the pectoral major muscle
that attaches to the antero-inferior surface of the clavicle. Its strong
constriction not only lowers the clavicle, but also rotates it anteriorly.
Final examination
While preparing the manuscript, the patient was once again
questioned for the circumstances of the incidence of destabilization.
When acquitted with presented above biomechanical analysis he
confessed that the plate broke, while fishing. Throwing a spinner for
six hours, once per every thirty seconds, he felt the pain at the clavicle.
Moreover, few days earlier he participated in a party drinking much
too much of alcohol that resulted in loose of his consciousness. As he
knows, any kind of additional trauma occurred during this episode,
but it could not be excluded, as some itching at the collar bone was
felt the following day.
Assuming a dynamic movements of the unloaded extremity
while fishing, that is subjected to the mass of a limb (ca 8 kg of the
weight and 66 cm distance from the acromion to the fist) loading
of the acromio–clavicular joint with the force of 80 N produces the
moment of the force of ca 26,4 Nm (80 N acting over a half of the
length of the extremity 330 mm). It is more than twice as much as
that sufficient to deform the plate permanently, as our experimental
bending test revealed that the plate already bends at the moment of
the force of 12,8 Nm. Moreover, dynamics of the movements while
throwing the spinner produces some additional moment of the force
that, depending on the position of the angler, significantly exceeds
those values. Those values are additionally increased by the weight of
the spinning rod and the spinner that, albeit relatively light (1,0 to 1,5
kg), but acting over the long arm, contribute to the destructive loads.
Figure 5
Figure 5
Inferior bending of the anterior and inferior edges of the anterior
branch of the plate (dashed lines). Inferior aspect of the plate.
Figure 6
Figure 6
Schematic representation of the process of plate fracture from the
superior (left) and postero-anterior views (right). Primarily, mechanical loads
acting on the distal part of the plate resulted in brittle fracture of its posterior
branch, than anterior one.
Discussion
Fracture destabilizations resulting from implant break offs are
common complications that require secondary interventions [7].
They encumber surgeons, hospitals and implants’ producers with
responsibility for unsatisfactory treatment bringing about the risk
of patient’s claims. Especially, when secondary injuries impair the
healing leading to non-union. But in some cases they may result
from intended patient’s activity or various unintended events in
close-minded individuals that contribute to the incident. Thus, the
question, which is in fact responsible for it, remains controversial.
Unfortunately, in most cases the real circumstances of the episode
remain unknown, posing patient’s secret.
Orthopedic implants are constructed to withstand mechanical
loads that are usually carried out by unloaded limb. But it does
not mean that they have to abide massive overloads. To reduce
disadvantageous mass of implanted metal, their dimensions are
optimized to transducer loads coming from the limb weight and
muscle tension of unloaded extremity only, with some safety margin,
basing on the idea that “the sufficient minimum is the best”. Thus,
their constructions do not allow massive strains bringing about the
risk of implant failure, when not abided. Titanium alloys, due to
modulus of elasticity comparable to that of a bone, are believed to be
the most suitable for plate manufacturing [8]. They were proved to
possess the strength sufficient to introduce an early rehabilitation [9].
In vast majority of cases patients obey postoperative
recommendations avoiding overloading of the stabilized extremity,
and orthotic immobilization for a sufficient period of time helps them
to avoid complications. But in some cases patients take off orthoses
and perform forbidden activities that result in implant failure or
losing. Especially, when they over use alcohol or drugs [10]. The loss
of self-control, irrational behavior and risky activities predispose
them to additional injuries, bringing about the risk of implant failure
[11]. Thus, in this particular group of patients additional protection
of fracture instrumentation has to be considered [12]. Even, when
alcohol overuses is occasional, as in the described above case.
Secondary injuries and inadequate limb immobilization may
complicate fracture stabilizations resulting in implant failure [13].
They usually affect lower extremity procedures, as those undergo
far bigger loads than upper ones. But, under certain circumstances,
breakings may occur in upper limbs. In analyzed case, repeated and
dynamic loads led to the break of the clavicle fracture stabilizing plate
clearly showing an undertaking of the physical efforts by the patient
disobeying postoperative recommendations.
In the case of clavicle stabilization, previously recommended for
a long (up to four weeks) period of time plaster casts are nowadays
usually reduced to maximum three weeks orthotic immobilizations,
as those were estimated to be sufficient, far more comfortable, and
allowing to avoid the risk of joint stiffness and muscular atrophy [14].
The effectiveness and usefulness of orthoses in postoperative fracture
management is widely accepted [15]. Nevertheless, the easiness to
take them off allow maneuvers overloading an implant and deprive
the surgeon’s control over the abide of recommendations. Thus,
those are risky in unreliable patients, as enable banned behavior. The
problem especially arises in mentally ill and alcohol and drug abused
persons, since behavior of those is highly unpredictable. But it is also
likely that some patients do not accept recommendations, or those
recommendations are simply unsuitable due to some reasons. Patients
could also provoke complications actively, as those allow them to
fix some vital problems (release from the jail, postpone of the case,
reprieve of the verdict, or simply obtain the compensation). But it
also happens that it is unintended and results from incomprehension
of the idea of those recommendations. Thus, profound explanation
of the aim of recommendations may improve patient’s cooperation.
Herein, we presented the patient, who informed about the
results of biomechanical analysis of the incidence confessed that
the implant broke while fishing, not primarily reported closure of
the car trunk lid. It points out to the fact that patients quite often
ignore recommendations and hide the information about the real
circumstances of the onset of the complication. Unfortunately, we
could not exclude that this confession is not true, as we do not have
an ability to verify it. It could be quite often observed that patients
verify the durability of stabilizations taking exercises, usually
young, physically active ones. Simple push-ups in an individual of
described patient’s posture load upper limbs with the weight of 40
kg to 45 kg. Analogically, other popular exercises including pull-ups,
weight lifting or working with resistance bands overload the limb
with forces that are sufficient to break the implant. Those episodes
are casuistically brought to date due, because patients do not report
those giving imaginary explanations. But insubordinate patients do
really exist bringing about problems not seen in regular ones. To
avoid those problems the best solution is to enforce them to follow
recommendations. In the case of bone fracture limb immobilization
in plaster cast is the most reliable solution, as it’s taking off is not easy
and every intervention is obvious and undeniable on examination,
giving an insight into patient’s contribution to the complication.
References
- Bostman O, Manninen M, Pihlajamaki H. Complications of plate fixation in fresh displaced mid clavicular fractures. J Trauma. 1997;43(5):778-83.
- Shin SJ, Do NH, Jang KY. Risk factors for postoperative complications of displaced clavicular midshaft fractures. J Trauma Acute Care Surg. 2012;72(4):1046-50.
- Peivandi MT, Yusof-Sani MR, Amel-Farzad H. Exploring the reasons for orthopedic implant failure in traumatic fractures of the lower limb. Arch Iran Med. 2013;16(8):478-82.
- Chen CE, Weng LH, Ko JY, Wang CJ. Management of nonunion associated with broken intramedullary nail of the femur. Orthopedics. 2008;31(1):78.
- Naimark M, Dufka FL, Han R, Sing DC, Toogood P, Ma CB, et al. Plate fixation of midshaft clavicular fractures: patient-reported outcomes and hardware-related complications. J Shoulder Elbow Surg. 2016;25(5):739-46.
- Jones GL, McCluskey GM 3rd, Curd DT. Nonunion of the fractured clavicle: Evaluation, etiology, and treatment. J South Orthop Assoc. 2000;9(1):43-54.
- Meeuwis MA, Pull Ter Gunne AF, Verhofstad MH, van der Heijden FH. Construct failure after open reduction and plate fixation of displaced midshaft clavicular fractures. Injury. 2017;48(3):715-9.
- Navarro M, Michiardi A, Castaño O, Planell JA. Biomaterials in orthopaedics. J R Soc Interface. 2008;5(27):1137-58.
- Goswami T, Markert RJ, Anderson CG, Sundaram SS, Crosby LA. Biomechanical evaluation of a pre-contoured clavicle plate. J Shoulder Elbow Surg. 2008;17(5):815-8.
- Schemitsch LA, Schemitsch EH, Kuzyk P, McKee MD. Prognostic Factors for Reoperation after Plate Fixation of the Midshaft Clavicle. J Orthop Trauma. 2015;29(12):533-7.
- Schemitsch LA, Schemitsch EH, Kuzyk P, McKee MD. Prognostic Factors for Reoperation after Plate Fixation of the Midshaft Clavicle. J Orthop Trauma. 2015;29(12):533-7.
- Berendes S, Zilkens C, Anastasiadis A, Graf M, Muhr G, Kalicke T. Additional external hinged fixator after open repositioning and internal fixation of acute elbow instability in non-compliant patients. Orthop Rev (Pavia). 2010;2(2):e21.
- Sharma AK, Kumar A, Joshi GR, John JT. Retrospective Study of Implant Failure in Orthopaedic Surgery. Med J Armed Forces India. 2006;62(1):70-2.
- Ersen E, Atalar AC, Birisik F, Saglam Y, Demirhan M. Comparison of simple arm sling and figure of eight clavicular bandage for midshaft clavicular fractures: A randomised controlled study. Bone Joint J. 2015;97-B(11):1562-5.
- Lusardi MM, Jorge M, Nielsen CC. Orthotics and prosthetics in rehabilitation. 3rd Ed. St. Louis: Elservier Saunders; 2013.