Case Report
Small Vessels, Big Solutions; Use of Graft Conduit to Facilitate Transfemoral Impella CP Implantation
Monteagudo Vela M*, Garcia Saez D, Zych B and Simon A
Department of Cardiothoracic Transplantation and Mechanical Circulatory Support, Harefield Hospital, Royal Brompton and Harefield NHS Foundation Trust, London, UK
*Corresponding author: Maria Monteagudo Vela, Department of Cardiothoracic Transplantation and Mechanical Circulatory Support, Harefield Hospital, Royal Brompton and Harefield NHS Foundation Trust, London, UK
Published: 12 Jul, 2018
Cite this article as: Monteagudo Vela M, Garcia Saez
D, Zych B, Simon A. Small Vessels,
Big Solutions; Use of Graft Conduit
to Facilitate Transfemoral Impella CP
Implantation. Clin Surg. 2018; 3: 2009.
Abstract
Mechanical circulatory support appeared to provide us with time; time for recovery, for decision
or for assessment. However they are known to be associated with different rates of mortality and
morbidity. Impella device is a less invasive mechanical circulatory support but provides similar
hemodynamic support compared with LVAD.
Advantages include respiratory wean and early mobilization, lower anticoagulation requirements
and less invasive surgical technique. Implanting options are limited and depend on the calibre of the
vessel. A Transaxillary approach need a minimum diameter of 6mm. Remaining options includes
transfemoral percutaneous insertion or a cut down femoral approach.
Keywords: Impella; Transfemoral approach; Cardiogenic shock
Introduction
Mechanical circulatory support has gained a main role in the heart failure pantheon. Impella is a less invasive mechanical assist device but is able to provide similar haemodynamic support as long term assist devices [1]. Advantages of Impella include early mobilisation and respiratory wean, as well as lower anticoagulation requirements and less invasive surgical techniques. Herein we present the case of a 16 year old teenager with known diagnosis of familiar dilated cardiomyopathy.
Case Presentation
He presented to our unit with decompensated heart failure and chest pain. Despite optimal
medical therapy including inotropic support he was in refractory cardiogenic shock. Therefore the
decision was made to escalate his therapy using the Impella CP mechanical assist device as a bridge
to heart transplantation.
Initially a transaxillary approach was attempted however this was abandoned due to small
caliber of the vessel (less than 6 mm which is the minimum diameter required). Remaining options
included a transfemoral percutaneous insertion or a cut down femoral approach. Given the small
yet adequate size of the femoral arteries we decided to perform a cut down femoral approach.
Dissection and exposure of the vessel was followed by the anastomosis of a 6 mm Dacron graft
that was tunnelled through a small incision in the upper thigh. After positioning the dedicated
0,025 inc wire into the left ventricle, the Impella was then railroaded into optimal position using
transoesophageal echocardiography. The 14 frenchpeel away sheath was not used in this case due
to the size of the graft. A minimum graft of 8mm is needed in order to introduce this sheath [2]
(Figure 1).
The device was secured and fixed to the skin. The wound was
sutured and the Impella was started achieving flows up to 3-3,5 l/min
[3] (Figure 2).
Figure 1
Figure 2
Conclusion
This is the first time this technique for implantation of an Impella CP has been described. In our experience, when Impella CP’s have been implanted using the percutaneous transfemoral approach as a bridge to long term mechanical assist devices or transplantation, we were faced with the issue of continuous oozing from the insertion site leading to the development of groin haematomas and significant haemoglobin drops over the course of few days. This occurred despite uncomplicated ultrasound guided puncture and fluoroscopic guided insertion of the Impella in the catheter laboratory. The proposed technique can be used for patients with borderline size vessels who required Impella support as a bridge to destination therapy.